|
|
|||||||||||||||||||||||||||||||||
| SECTION
1 Start Up
SECTION 2 Zooming In
SECTION 3
|
Start
Up Procedure:
Parts of the microscope system. 1. Turn on mercury light. (rocker switch-ON, press and hold IGNITE button until the “click” is heard). 2. Start computer. Log into Windows NT. Do not start Lasersharp yet.
Preparing to Image the Specimen
9. Start laser with blue “star” button near top of the scanner control pane. 10. Optimize Brightness and Contrast A. Adjust laser, gain and iris until brightness is optimized and the iris is set as close to optimum as possible (click iris icon on right side of screen to set optimally). · Do not close down iris unless increasing the laser power or gain will compensate for decreased brightness! · If more brightness is needed, try increasing laser power (this is a very low power laser and will NOT bleach the specimen rapidly). · Try to keep gain set below 33%. At 33 and above noise increases exponentially. B. Adjust brightness settings (with setcol loaded) until most of background is green (brightness level of zero) and structures in focus are mostly gray, with a few pixels of red (brightness = saturated (level of 256)).
11.
Determine level of signal averaging appropriate to sample. A. Set Kalman in “collection filter” B. Leave n = stop (Ignore the Factor setting, it has no function for this filter) C. Start laser and watch the image, count passes as the screen updates. D. Stop laser when the image has smoothed sufficiently. E. Note the value found in step C, it will be needed when collecting a stack (below).
1. Set stack start and stop positions. A. Set collection filter to direct (scanner control pane). B. Start laser.
· In the position box, click on the down arrow until desired location is reached. · Mark position by clicking on start (in limits section). D. Find stop position: · Click on up arrow until at desired location. · Mark position by clicking on stop. E. Stop laser. F. Set Kalman = n, where n = the value found in step 11.c in the above section.
3. In the pane that opens, select enable z series (!) 4. Click on “start” and your stack will be collected.
Note: Limit zooms to central regions of the field. If an object you want is not centered, use the stage controls to center it.
A technique that prevents "bleed-through" from G to R: 1. Adjust brightness for each channel individually in channels box. A. Choose one of the individual channel panes you want to collect. B. Adjust iris, gain, black level and laser power C. Choose next channel, adjust, etc. 2.
Begin collection by clicking on the stack
icon. 3.
Click on the channels
tab. 4.
Select sequential. 5. Click “start”.
1. If not already opened, open the image file. 2. The collected stack can be viewed manually or automatically. - To view automatically, click on the “back & forth” icon at top of image box. 3. Examine the data (make projections, rotations, add scale bars, etc).
1. Exit LaserSharp by selecting file, then logout. 2. Turn off laser power supply (turn key CCW) 3. Turn off Hg-vapor power supply. 4. Turn off control box (under table) 5. Data is accessible through two methods: via FTP, or by burning a CD-ROM. 6. When done copying files, logout from NT and turn off monitor. Please leave the computer running.
Software
that
opens BioRad “PIC” files: Note: A BioRad PIC file is not a Macintosh PIC file!
|
||||||||||||||||||||||||||||||||
|
This page maintained by bcoats@seas.upenn.edu Copyright © 2000 Department of
Bioengineering - University of Pennsylvania. All rights reserved.
|